13-25 Seal Integrity Testing Devices

Seal Integrity Testing Devices photo.

INTRODUCTION

These products are used to test the seal integrity of a flexible package. The creep test function checks the package seal under sustained pressure. The creep-toburst test function sequentially checks minimum required seal strength and maximum pressure at burst. Tests can also simulate high altitude conditions.

APPLICATIONS

Flexible package testing for Sterilized Medical Devices,
Food and Pharmaceuticals.

SPECIFICATIONS

  • Model 2600 Seal Integrity Tester checks the seal integrity on a wide range of porous and non-porous flexible packages using a pressurized air burst test. Open and closed package test fixtures are available in a variety of standard and custom configurations. Pressurized burst test uses ASTM 1140 method.
  • Model 2700 Integrity Tester detects leaks in non-porous packages with a vacuum test chamber. Leak location can be identified by submerging the sample in water.

Seal Integrity Testing Devices Sensing Probes photo.

FEATURES

Model 2600 Seal Strength Tester

  • RS 232 data acquisition interface
  • Automatic Pressure Control
  • 0.2% full scale pressure calibration accuracy

CONNECTIONS

Electrical: Specify voltage requirements when ordering

PACKAGE TEST SYSTEM SPECIFICATIONS

Vacuum Source

Use vacuum pump 13-26-01 or equivalent

Electrical Power

85-264 VAC 47-63 Hz

Mechanical

Weight:
Control Console: 11.4 lbs (5.2 kg)
Vacuum Chamber: Approximately 3 lbs (1.4 kg)
Parallel Printer Port Standard db 25 female connector, Centronics format

Environmental

Temperature Range
Operating: 40°F - 110°F (4°C - 43°C)
Non-operating: 0°F - 125°F (-184°C - 52 °C)
Humidity: Up to 90% R.H.

Preformance

Vacuum Accuracy 0.2% full scale 29.9 in. Hg
Seal Integrity Testing Devices photo.
Seal Integrity Testing Devices photo.
Seal Integrity Testing Devices photo.
Testing Range mmHg lnHg Altitude
Absolute Pressure (max) 760 29.9 N/A
Absolute Pressure (min) 1 1 N/A
Gauge Pressure (max) 760 29.9 N/A
Gauge Pressure (min) 1 1 N/A
Altitude (max) N/A N/A 50,000
Altitude (min) N/A N/A depends on geographic location

OPEN PACKAGE TEST FIXTURES

13-25 Seal Integrity Testing Devices photo
13-25 Seal Integrity Testing Devices photo

AUTOMATIC CONTROL CONSOLE

Vacuum Source

70-100 PSIG Dry, oil-free, instrument quality air

Electrical Power

85-264 VAC 47-63 Hz

Mechanical

Weight: 13 lbs (5.9 kg)
Serial (RS232) Output Port: Standard DB 9 female connector
Parallel Printer Port Standard db 25 female connector, Centronics format

Environmental

Temperature Range
Operating: 40°F - 110°F (4°C - 43°C)
Non-operating: 0°F - 125°F (-184°C - 52 °C)
Humidity: Up to 90% R.H.

Preformance

Pressure Accuracy 0.2% full scale (50 PSI)
Pressure Resolution 0.3" H20
Sensitivity Adjustments 9 Programmable Settings
Flow Rate Adjustments 9 Programmable Settings
Seal Integrity Testing Devices photo.
Seal Integrity Testing Devices photo.
Seal Integrity Testing Devices photo.

ORDERING INFORMATION

Model 2600 Catalog Number 13-25
Model 2700 Catalog Number 13-26

Sensing Probes:

Part No. Description Weight
13-25-05 Closed Package Test Fixture with 3/16 in. Diameter Needle Probe Assembly 2.7 kg (6.0 lbs.)
13-25-07 Closed Package Test Fixture with Double Needle Probe Assembly  

Open Package Test Fixture:

Part No. Description Weight
13-25-01 12” Open Package Test Fixture 3.2 kg (7 lbs.)
13-25-02 24” Open Package Test Fixture 10.4 kg (23 lbs.)

Automatic Control Console:

Part No. Power Supply Printer Output RS232 Output Port Data Export
13-25-00-0001 110 VAC; 60 Hz Yes Yes Yes
13-25-00-0002 220 VAC; 50 Hz Yes Yes Yes

Call Us Today

For additional information on Shows and Expos, or to learn how Testing Machines can help your company profit through testing, please call us at 1-800-678-3221, or e-mail info@testingmachines.com.

spacer

Search Testing Machines:

 

PDF Product Sheet:

Available in PDF format. Seal Integrity Testing Devices

Accessories:

Closed Package Test Fixture provides a stable platform to perform pressurized seal strength tests on closed, flexible packages.
Open Package Test Fixtures provide convenient clamping of a package sample during pressurized seal testing.

The TMI Catalog on CD-ROM:

The TMI Group of Companies Product Catalog on CD-ROM photo.

The latest catalog of physical property testing equipment from the TMI Group of Companies is now available on CD-ROM.

Click here to request a copy or call 631-439-5400 or 1-800-678-3221 (from with U.S. and Canada).

The TMI Guide to Pulp and
Paper Testing:

Available in PDF format.Click here for the TMI Guide to Pulp and Paper Testing in PDF format.

Click here to download Adobe Acrobat Reader. It's free.

©2007 Testing Machines Inc. All rights reserved.